Investigation of permanent magnets in low-cost position tracking

TitleInvestigation of permanent magnets in low-cost position tracking
Publication TypeConference Paper
Year of Publication2016
AuthorsAnderson, R., Lasso A., Hashtrudi-Zaad K., & Fichtinger G.
Conference NameProc. SPIE
Pagination97862F-97862F-9
Abstract

PURPOSE: Low cost portable ultrasound systems could see improved utility if similarly low cost portable trackers were developed. Permanent magnet based tracking systems potentially offer adequate tracking accuracy in a small workspace suitable for ultrasound image reconstruction. In this study the use of simple permanent magnet tracking techniques is investigated to determine feasibility for use in an ultrasound tracking system. METHODS: Permanent magnet tracking requires finding a position input into a field model which minimizes the error between the measured field, and the field expected from the model. A simulator was developed in MATLAB to determine the effect of sources of error in permanent magnet tracking systems. Insights from the simulations were used to develop a calibration and tracking experiment to determine the accuracy of a simple and low cost permanent magnet tracking system. RESULTS: Simulation and experimental results show permanent magnet based tracking to be highly sensitive to errors in sensor measurements, calibration and experimental setup. The reduction in field strength of permanent magnets lowers with the cube of distance, which leads to very poor signal-to-noise ratios at distances above 20 cm. Small errors in experimental setup also led to high tracking error. CONCLUSION: Permanent magnet tracking was found to be less accurate than is clinically useful, and highly sensitive to errors in sensors and experimental setup. Sensor and calibration limitations make simple permanent magnet tracking systems a poor choice given the present state of sensor technology.

URLhttp://dx.doi.org/10.1117/12.2217189
DOI10.1117/12.2217189
PerkWeb Citation Keydoi:10.1117/12.2217189